4.8.4. ANALOG I/O CONFIGURATION
Table 6-8 lists the analog I/O functions that are available in the T100.
Table 4-9: DIAG - Analog I/O Functions
SUB MENU
FUNCTION
AOUTS CALIBRATED:
Shows the status of the analog output calibration (YES/NO) and initiates a calibration of all
analog output channels.
CONC_OUT_1
Sets the basic electronic configuration of the A1 analog output (SO
2
). There are three options:
RANGE:
Selects the signal type (voltage or current loop) and full scale level of the
output.
REC_OFS:
Allows setting a voltage offset (not available when RANGE is set to Current
Loop (
CURR
).
AUTO_CAL:
Performs the same calibration as AOUT CALIBRATED, but on this one
channel only.
NOTE: Any change to RANGE or REC_OFS requires recalibration of this output.
CONC_OUT_2
Same as for CONC_OUT_1 but for analog channel 2 (SO
2
)
TEST OUTPUT
Same as for CONC_OUT_1 but for analog channel 3 (TEST)
CONC_OUT_3
(Not a
vailable in the analyzer’s standard configuration; applies when optional sensor installed).
AIN CALIBRATED
Shows the calibration status (YES/NO) and initiates a calibration of the analog input channels.
XIN1
.
.
.
XIN8
For each of 8 external analog inputs channels, shows the gain, offset, engineering units, and
whether the channel is to show up as a Test function.
Table 4-10: Analog Output Voltage Ranges
RANGE
MINIMUM OUTPUT
MAXIMUM OUTPUT
0-0.1 V
-5 mV
+105 mV
0-1 V
-0.05 V
+1.05 V
0-5 V
-0.25 V
+5.25 V
0-10 V
-0.5 V
+10.5 V
The default offset for all ranges is 0 VDC.
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...