Operating Instructions
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
98
4.6.8. DILUTION RATIO (OPTION)
The dilution ratio is a software option that allows the user to compensate for any dilution
of the sample gas before it enters the sample inlet. Once the degree of dilution is known,
add an appropriate scaling factor to the analyzer’s SO2 concentration calculation so that
the measurement range and concentration values reflect the undiluted values when
shown on the instrument’s front panel display screen and reported via the analog and
serial outputs.
Using the dilution ratio option is a 4-step process:
1. Select reporting range units: Follow the procedure in Section 6.7.7.? 6.7.8?
2. Select the range: Use the procedures in Section 6.7.4
– 6.7.6. Ensure that the
SPAN value entered is the maximum expected concentration of the undiluted
calibration gas and that the span gas is either supplied through the same dilution
inlet system as the sample gas or has an appropriately lower actual concentration.
For example, with a dilution set to 100, a 1 ppm gas can be used to calibrate a 100
ppm sample gas if the span gas is not routed through the dilution system. On the
other hand, if a 100 ppm span gas is used, it needs to pass through the same
dilution steps as the sample gas.
3. Set the dilution factor as a gain (e.g., a value of 20 means 20 parts diluent and 1
part of sample gas):
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
SETUP C.3
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP C.3
RANGE CONTROL MENU
MODE
SET UNIT
DIL
EXIT
DIL
only appears
if the dilution ratio
option has been
installed
SETUP C.3
DIL FACTOR: 1.0 GAIN
0 0 0 1 .0
ENTR
EXIT
Toggle these keys to set the dilution
factor.
This is the number by which the
analyzer will multiply the SO
2
concentrations of the gas passing
through the reaction cell.
SETUP C.3
DIL FACTOR: 20.0 GAIN
0 0 2
0 .0
ENTR
EXIT
EXIT
ignores the
new setting.
ENTR
accepts the
new setting.
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
Figure 4-15:
SETUP RNGE
– Dilution Ratio
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...