Theory Of Operation
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
232
06807A(DCNxxxx)
Figure 9-7: Oxygen Sensor - Theory of Operation
9.2.2. O
2
SENSOR OPERATION WITHIN THE T100 ANALYZER
The oxygen sensor option is transparently integrated into the core analyzer operation.
All functions can be viewed or accessed through the front panel display, just like the
functions for SO
2
.
The O
2
concentration is displayed below the SO
2
concentration.
Test functions for O
2
slope and offset are viewable from the front panel along with the other
test functions of the analyzer.
O
2
sensor calibration is performed via the front panel
CAL
function and is performed in a
nearly identical manner as the standard SO
2
calibration. Refer to Section 6.10.1 for more
details.
Stability of the O
2
sensor can be viewed via the front panel (refer to Section 6.10.1.3).
The O
2
concentration range is 0-100% (user selectable) with 0.1% precision and
accuracy.
The temperature of the O
2
sensor is maintained at a constant 50°C by means of a PID
loop and can be viewed on the front panel as test function O
2
TEMP.
The O
2
sensor assembly itself does not have any serviceable parts and is enclosed in an
insulated canister.
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...