Teledyne ML
- T100 UV Fluorescence SO2 Analyzer
EPA Protocol Calibration
205
Wait until
STABIL
falls
below
0.5 ppb.
This may take
several minutes.
SAMPLE
STABIL=X.XXX PPB
SO2 =XXX.X
< TST TST > CAL
CALZ
CALS
SETUP
Record the SO
2
reading as
displayed on the
instrument’s front
panel
ACTION:
Allow calibration gas diluted to proper concentration for
Midpoint N to enter the sample port
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST
TST
>
CAL
SETUP
SAMPLE
STABIL=X.XXX PPB
SO2 =XXX.X
< TST TST >
CAL
SETUP
Set the Display to show the
STABIL
test function.
This function calculates the
stability of the SO
2
measurement
SPAN CAL M
RANGE = 500.0 PPB
SO
2
XXX.X
< TST TST >
ZERO
SPAN
CONC
EXIT
ACTION:
Allow Calibration Gas diluted to
proper concentration for
Midpoint N+1 to enter the sample
port
Press
EXIT
to
Return to the
Main
SAMPLE
Display
Figure 7-1: Dynamic Multipoint Span Calibration
7.6. SPECIAL CALIBRATION REQUIREMENTS FOR DUAL RANGE
OR AUTO RANGE
If Dual Range or Auto Range is selected, then calibration for Range1 and Range2
should be performed separately.
For zero and span point calibration, follow the procedure described in Section 6.2.
Repeat the procedure for both the
HIGH
and
LOW
Ranges.
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...