125
5.
REMOTE OPERATION & ADVANCED FEATURES
5.1. SETUP
– COMM: SETTING UP THE ANALYZER’S
COMMUNICATION PORTS
For remote operation the T100 is equipped with an Ethernet port, a USB port and two
serial communication (COMM) ports located on the rear panel (refer to Figure 3-4).
Both COMM ports (labeled
RS232
, which is the COM1 port, and
COM2
) operate
similarly and give the user the ability to communicate with, issue commands to, and
receive data from the analyzer through an external computer system or terminal. By
default, both ports operate on the RS-232 protocol.
The
RS232
port can also be configured to operate in single or RS-232 Multidrop mode
(Option 62); refer to Sections 3.4.8.3 and 3.4.8.4 respectively.
The
COM2
port can be left in its default configuration for standard RS-232 operation,
or reconfigured for half-duplex RS-485 operation; (Section 3.4.8.5). When
COM2
is
configured for RS-485 communication, the rear panel USB port is disabled.
A code-activated switch (CAS), can also be used on either port to connect typically
between 2 and 16 send/receive instruments (host computer(s) printers, data loggers,
analyzers, monitors, calibrators, etc.) into one communications hub. Contact Teledyne
ML’s sales for more information on CAS systems.
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...