4.6.3. REPORTING RANGE MODES
The T100 provides three analog output range modes to choose from:
Single range (
SNGL
)
mode sets a single maximum range for the analog output. If single
range is selected (refer to Section 4.6.4) both outputs are slaved together and will represent
the same measurement span (e.g. 0-50 ppm), however their electronic signal levels may be
configured for different ranges (e.g. 0-10 VDC vs. 0-.1 VDC – Refer to Section 4.8.4.1).
Dual range (
DUAL
) allows the A1 and A2 outputs to be configured with different
measurement spans (refer to Section 4.6.5).
Auto range (
AUTO
) mode gives the analyzer to ability to output data via a low range and
high range. When this mode is selected (refer to Section 4.6.6) the T100 will automatically
switch between the two ranges dynamically as the concentration value fluctuates.
Also, in this mode the RANGE Test function displayed on the front panel during
SAMPLE mode will be replaced by two separate functions, Range1 and Range2.
Range status is also output via the External Digital I/O Status Bits (refer to Section
5.3.1.1).
To select the Analog Output Range Type press:
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP X.X
RANGE MODE:
SNGL
SNGL
DUAL AUTO
ENTR EXIT
EXIT
Returns
to the Main
SAMPLE
Display
Go To
Section
6.7.4
Go To
Section
6.7.5
Go To
Section
6.7.6
Only one of the
range modes may
be active at any
time.
Figure 4-10:
SETUP RNGE
– Reporting Range Mode
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...