
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Operating Instructions
95
4.6.5. DUAL RANGE MODE (DUAL)
Selecting Dual Range mode allows the
A1
and
A2
outputs to be configured with
different reporting ranges. The analyzer software calls these two ranges low and high.
The Low range setting corresponds with the analog output labeled
A1
on the rear panel
of the instrument. The high range setting corresponds with the
A2
output. While the
software names these two ranges low and high, they do not have to be configured that
way. For example: the low range can be set for a span of 0-150 ppb while the high range
is set for 0-50 ppb.
In
DUAL
range mode the
RANGE
test function displayed on the front panel will be
replaced by two separate functions:
range1: The range setting for the A1 output.
range2: The range setting for the A2 output.
To set the ranges press following control button sequence
.
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP X.X
RANGE MODE:
SNGL
SNGL
DUAL
AUTO
ENTR EXIT
SETUP X.X
RANGE MODE:
DUAL
SNGL
DUAL AUTO
ENTR
EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP X.X
HIGH RANGE:
500.0 Conc
0 0 5
0 0 .0
ENTR
EXIT
SETUP X.X
LOW RANGE
:
500.0 Conc
0 0
1
0 0 .0
ENTR
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
EXIT
Returns
to the Main
SAMPLE
Display
Toggle the
numeral buttons
to set the upper
limit of each
range.
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
Figure 4-12:
SETUP RNGE
– Dual Range Mode
IMPORTANT
In DUAL range mode the LOW and HIGH ranges have separate slopes and offsets for
computing SO
2
concentration. The two ranges must be independently calibrated.
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...