Teledyne ML - T100 UV Fluorescence SO2 Analyzer
APPENDIX A - Version Specific Software Documentation
A-41
APPENDIX A-8: Setting up Communications
Setting up the communications for integration with a Teledyne Monitor Labs system.
The following describes the settings that should be entered into the analyzer’s COMM sub menus.
Ethernet Communication
The following settings must be entered into the comm. menus in order to achieve Ethernet
communication. Analyzers shipped with systems are setup at the factory.
ID: Analyzer specific. See table below.
INET:
DHCP: OFF
INST IP: site specific
GATEWAY IP: leave as factory default or site specific.
SUBNET MASK: leave as factory default or site specific.
TCP PORT: 3000
ONLINE: ON
BAUD RATE: 115200
Analyzer ID Table
SO
2
TRS
NO
x
CO
CO
2
O2
O
2
/CO
2
MACHINE
ID
100 102 200 300 360 802
GAS
ID 150 160 141 130 120 170
Serial Communication
Teledyne Monitor Labs systems communicate to the analyzers via the PLC using serial communication.
The communication protocol used in the analyzer is the Hessen two protocol. When an analyzer is first
turned on, the Hessen variables do not exist in the non-volatile memory. In order to create these
variables, the Hessen protocol must be turned on in COM 1 and then the power must be cycled.
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...