9. HVPS fine adjustment switch (Range 0-9, then A-F).
10. Gain adjustment potentiometer (Full scale is 10 to 12 turns).
11. Set the HVPS coarse adjustment to its minimum setting (0).
12. Set the HVPS fine adjustment switch to its maximum setting (F).
13. Turn the gain adjustment potentiometer clockwise to its maximum setting.
14. Set the front panel display to show
STABIL
(refer to Section 4.2.1).
15. Feed span gas into the analyzer.
16. Wait until the
STABIL
value is below 0.5 ppb.
IMPORTANT
Use a span gas equal to 80% of the reporting range.
Example: for a reporting range of 500 ppb, use a span gas of 400 ppb.
17. Scroll to the
OFFSET
function and record the value.
18. Scroll to the
NORM PMT
value.
ATTENTION
COULD DAMAGE INSTRUMENT
Do not overload the PMT by accidentally setting both adjustment switches to their
maximum setting. This can cause permanent damage to the PMT.
19. Determine the target
NORM PMT
value according to the following formulas.
If the reporting range is set for
2,000 ppb (the instrument will be using the
2,000 ppb physical range):
Target
NORM PMT
=
(2 x span gas concentration) +
OFFSET
If the reporting range is set for
2,001 ppb(the instrument will be using the
20,000 ppb physical range):
Target
NORM PMT
=
(0.181 x span gas concentration) +
OFFSET
EXAMPLE
: If the
OFFSET IS
33 mV, the Reporting Range is 500 ppb, the span
gas should be 400 ppb and the calculation would be:
Target
NORM PMT
=(2 x 400) + 33 mV
Target
NORM PMT
=
833 mV
20. Set the HVPS coarse adjustment switch to the lowest setting that will give you more
than the target NORM PMT signal from Step 16.
The coarse adjustment typically increments the
NORM PMT
signal in 100-300
mV steps.
21. Adjust the HVPS fine adjustment such that the NORM PMT value is at or just above
the target NORM PMT signal from Step 16.
22. Continue adjusting the both the coarse and fine switches until norm PMT is as close
to (but not below) the target NORM PMT signal from Step 16.
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...