Remote Operation & Advanced Features
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
132
5.1.3. USB CONFIGURATION
After connecting a USB cable between your PC and the instrument, ensure their baud
rates match (change the baud rate setting for either your PC’s software or the
instrument). Section 5.1.5 shows how to navigate to the baud rate setting in the
Communications Menu where COM2 is the default setup menu for USB configuration.
Also, while there are various communication modes available (Table 5-4), the default
settings are recommended for USB, except to change the baud rate if desired.
5.1.4. COMM PORT COMMUNICATION MODES
Each of the analyzer’s serial ports can be configured to operate in a number of different
modes, which are listed in the following table. Each COMM port needs to be configured
independently. Refer to Appendix A for menu trees.
Table 5-4: COMM Port Communication modes
MODE
1
ID
DESCRIPTION
QUIET
1
Quiet mode suppresses any feedback from the analyzer (DAS reports, and warning
messages) to the remote device and is typically used when the port is communicating
with a computer program such as APICOM. Such feedback is still available but a
command must be issued to receive them.
COMPUTER
2
Computer mode inhibits echoing of typed characters and is used when the port is
communicating with a computer program, such as APICOM.
HESSEN
PROTOCOL
16
The Hessen communications protocol is used in some European countries. Teledyne
ML
’s part number 02252 contains more information on this protocol.
E, 7, 1
2048
When turned on this mode switches the
COMM
port settings
from
No parity; 8 data bits; 1 stop bit
to
Even parity; 7 data bits; 1 stop bit
RS-485
1024
Configures the
COM2
Port for RS-485 communication. RS-485 mode has precedence
over Multidrop mode if both are enabled. When the COM2 port is configured for RS-485
communication, the rear panel USB port is disabled.
SECURITY
4
When enabled, the serial port requires a password before it will respond. The only
command that is active is the help screen (? CR).
MULTIDROP
PROTOCOL
32
Multidrop protocol allows a multi-instrument configuration on a single communications
channel. Multidrop requires the use of instrument IDs.
ENABLE
MODEM
64
Enables sending a modem initialization string at power-up. Asserts certain lines in the
RS-232 port to enable the modem to communicate.
ERROR
CHECKING
2
128
Fixes certain types of parity errors at certain Hessen protocol installations.
XON/XOFF
HANDSHAKE
2
256
Disables XON/XOFF data flow control also known as software handshaking.
HARDWARE
HANDSHAKE
8
Enables CTS/RTS style hardwired transmission handshaking. This style of data
transmission handshaking is commonly used with modems or terminal emulation
protocols as well as by Teledyne Instrument’s APICOM software.
HARDWARE
FIFO
2
512
Improves data transfer rate when one of the COMM ports.
COMMAND
PROMPT
4096
Enables a command prompt when in terminal mode.
1
Modes are listed in the order in which they appear in the
SETUP
MORE
COMM
COM[1 OR 2]
MODE
menu
2
The default setting for this feature is
ON.
Do not disable unless instructed to by Teledyne ML
’s Customer Service personnel.
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...