Remote Operation & Advanced Features
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
126
SETUP X.X
COMMUNICATIONS MENU
ID
HESN
COM1
COM2
EXIT
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS DIAG
EXIT
SETUP X.X
COMMUNICATIONS MENU
ID
INET
COM1 COM2
EXIT
Menu with
Internet
Properties
Menu with
Hessen Protocol
Figure 5-1: SETUP
– COMM Menu
5.1.1. MACHINE (ANALYZER) ID
Each type of Teledyne ML analyzer is configured with a default ID code. The default
ID code for all T100 analyzers is
100
. The ID number is only important if more than
one analyzer is connected to the same communications channel such as when several
analyzers are on the same Ethernet LAN (refer to Section 5.1.2); in a RS-232 Multidrop
chain (refer to Section 3.4.8.4) or operating over a RS-485 network (refer to Section
3.4.8.5). If two analyzers of the same model type are used on one channel, the ID codes
of one or both of the instruments needs to be changed so that they are unique to the
instruments
To edit the instrument’s ID code, access the Machine ID from the COMM Menu (refer
to Figure 5-1), then press:
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...