3.5.6. PNEUMATIC LAYOUT WITH CO
2
SENSOR OPTION
Figure 3-22 shows the internal, pneumatic connections for the analyzer with the carbon
dioxide (CO
2
) sensor option installed. Pneumatically, the CO
2
sensor is placed in line
with the sample gas line between the particulate filter and the analyzer’s sample
chamber. It does not alter the gas flow rate of the sample through the analyzer.
FLOW PRESSURE
SENSOR PCA
Chassis
with CO
2
Sensor Option
EXHAUST
GAS OUTLET
UV
LAMP
PMT
REACTION
CELL
PUMP
Particulate
Filter
HYDROCARBON
SCRUBBER
(Kicker)
V
A
C
U
U
M
M
A
N
IF
O
L
D
CO
2
Probe
SAMPLE
GAS INLET
SAMPLE
PRESSURE
SENSOR
FLOW
SENSOR
Figure 3-22:
Pneumatic Layout with CO
2
Sensor
3.5.7. CALIBRATION GASES
Zero air and span gas are required for accurate calibration.
3.5.7.1.
ZERO AIR
A gas that is similar in chemical composition to the earth’s atmosphere but without the
gas being measured by the analyzer, in this case SO
2
. If your analyzer is equipped with
an Internal Zero Span (IZS) or an external zero air scrubber option, it is capable of
creating zero air.
For analyzers without an IZS or external zero air scrubber option, a zero air generator
such as the Teledyne Model 701 can be used (Figure 3-16).
3.5.7.2.
SPAN GAS
A gas specifically mixed to match the chemical composition of the type of gas being
measured at near full scale of the desired measurement range. In this case, SO
2
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...