3.
Before starting the procedure, the user must verify that the parameter values
are correct.
Check the parameter values for ‘Flap thickness [µm]:’ , ‘Diameter [mm]:’ ,
‘Bed energy [µJ]:’ and ‘Rim energy [µJ]:’ in the ‘Parameters’ dialog box
(Fig. 161), and check the position of the flap. Change it, if necessary.
4.
If you agree with the parameter setting, continue with step 7.
5.
If the parameter setting is not correct, either click the
‘Center to pupil’
button (Fig. 161) to center the flap automatically based on an automatical
pupil detection, or click within the red highlighted circle and shift it by
dragging and dropping in the right position to center the orange marking
circle manually.
Fig. 162: Centering is performed
ð
The flap is correctly positioned (Fig. 162).
L
WARNING! Risk of rough beds or tissue bridges due to move‐
ment when laser is firing!
Rough beds or tissue bridges may occur if vibrations from move‐
ment occur when the laser is firing.
– Do not walk around while the procedure is performed.
– Do not touch the patient bed or the patient's head while the
procedure is performed.
– Do not touch the suction clip.
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 52
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 178 of 207
Содержание Victus
Страница 1: ...User Manual VICTUS Femtosecond Laser Platform SW V3 3 SP02 UM 100010770 GB SKU 70005780TPV Version 6 ...
Страница 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Страница 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Страница 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Страница 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Страница 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...
Страница 202: ...Table of contents A EMC declaration APPENDIX 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 2 ...