2.3
Important safety considerations
Danger of injury due to user negligence!
Always double-check the following before treatment:
– Are you treating the right patient?
– Did you select the correct treatment?
– Was the correct eye selected?
– Does the entered data match the patient to be treated?
Failure to verify any of the above-mentioned aspects may result in serious
permanent patient injury.
Danger of injury for persons with cardiac pacemakers!
Negative influence on the pacemaker, such as electromagnetic interference,
could occur during laser operation.
– During laser operation, no person— medical personnel or patients—
with a cardiac pacemaker should be present in the treatment room.
Property damage due to improper stacking or locating of non-intended
devices next to the product!
The product is very sensitive. Improper stacking or locating of non-intended
devices next to the product may lead to equipment damage or influence
concerning EMC. If adjacent or stacked use is necessary, the product should
be observed to verify normal operation in the configuration in which it will
be used.
– Do not stack or locate any non-intended devices next to the product
during operation.
The
medical electrical equipment
implies special precautions regarding EMC and
needs to be installed and put into service according to the EMC information provided
in
Ä
Appendix A ‘EMC declaration’ on page 13 - 3.
The
medical electrical equipment
may be affected by portable and mobile radio fre‐
quency communication equipment.
2.4
Responsible organization
The responsible organization can be, for example, a public hospital, a private
person (ophtalmologist), or a private clinic.
Education and training is included in the term 'use'.
n
The responsible organization must ensure that hazards of network coupling
of the device must be observed. Such hazards include, for example, virus pro‐
tection, data security, network protection by firewall systems, and ensuring
sufficient bandwidth.
n
System modifications and use in combination with other equipment is
limited to TPV-approved equipment. Otherwise, the responsible organization
is in charge of proper risk management.
L
WARNING
L
WARNING
L
WARNING
SAFETY
2
SKU: 70005780TPV * Version 6 * UM-100010770 GB
2 - 3
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 19 of 207
Содержание Victus
Страница 1: ...User Manual VICTUS Femtosecond Laser Platform SW V3 3 SP02 UM 100010770 GB SKU 70005780TPV Version 6 ...
Страница 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Страница 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Страница 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Страница 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Страница 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...
Страница 202: ...Table of contents A EMC declaration APPENDIX 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 2 ...