9.4.2.1
Using a headrest with integrated joystick
1.
Press the red pushbutton (Fig. 111/2 or Fig. 112/3) on the side of the head‐
rest and allow 3 seconds for the lock to engage and disengage to facilitate
the smooth movement of the patient bed. Swing the patient bed out com‐
pletely.
Refer to the user manual "Laser Patient Bed LS Comfort" or "Patient Support
System PSS S60" by AKRUS GmbH & Co KG for instructions on how to operate
the patient bed.
2.
Ask the patient to lie down and ensure that the legs of the patient are not
crossed.
The supplied leg bolster should always be used to reduce patient movement.
3.
Place the leg bolster under the patient's knees to improve patient comfort
and discourage crossing of legs.
4.
Ask the patient to lay the arms alongside his body or fold them lightly
clasped across the stomach.
5.
Use the molded headrest to securely and comfortably fixate the patient's
head.
ð
The patient is positioned on the patient bed so that the cornea is in a
horizontal plane.
Fig. 111: Headrest with integrated joy‐
stick (
‘LS Comfort’
)
Fig. 112: Headrest with integrated joy‐
stick (
‘Patient Support System S60’
)
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 10
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 136 of 207
Содержание Victus
Страница 1: ...User Manual VICTUS Femtosecond Laser Platform SW V3 3 SP02 UM 100010770 GB SKU 70005780TPV Version 6 ...
Страница 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Страница 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Страница 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Страница 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Страница 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...
Страница 202: ...Table of contents A EMC declaration APPENDIX 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 2 ...