4.
Click the
‘OK’
button to close the message ‘Vacuum test successful.’ .
ð
The ‘Vacuum test’ screen closes and the status indicator turns green
(Fig. 104).
5.
Remove the suction reservoir dummy.
9.3.4
Performing the pressure test
Special tool:
n
Digital scale
n
Round plate holder
n
Short adapter
n
Support block
1.
Click the
‘System test’
button (Fig. 27/5) on the ‘Main menu’ screen.
2.
Click the
‘Pressure test’
button (Fig. 45/3) to open the ‘Pressure test’ screen.
Fig. 105:
‘Pressure test’
‘Preparation’
screen
3.
Follow the instructions in the ‘Pressure test’ screen.
Fig. 104:
‘Vacuum test’
successful
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 5
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 131 of 207