‘Laser Stop’ button with integrated ‘Bed Stop’
Action
Results
To activate the ‘Laser Stop’
button:
n
Press the ‘Laser Stop’
button .
Any possible laser emission stops immediately by cut‐
ting off the laser interlock.
The patient bed cannot be moved upwards or down‐
wards.
The integrated illumination in the ‘Laser Stop’ button is
no longer illuminated.
The ‘Reset’ button is illuminated simultaneousely.
To unlock the ‘Laser Stop’
button:
n
Pull the ‘Laser Stop’
button towards you so
that a restart is possible.
n
Press the ‘Reset’ button
(Fig. 9/5).
After 2 seconds the ‘Reset’ button is no longer illumi‐
nated (normal operation).
The integrated illumination in the ‘Laser Stop’ button is
illuminated red (normal operation).
Depending on the
headrest
connected to the patient bed, please note the following
warning!
Headrest with external joysticks
Headrests with integrated joysticks
Red pushbutton at the headrest (at
the patient bed ‘LS Comfort’
Pushbutton at the headrest (at the
patient bed ‘Patient Support System
S60’ )
Patient beds with headrests
SAFETY
2
SKU: 70005780TPV * Version 6 * UM-100010770 GB
2 - 10
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 26 of 207
Содержание Victus
Страница 1: ...User Manual VICTUS Femtosecond Laser Platform SW V3 3 SP02 UM 100010770 GB SKU 70005780TPV Version 6 ...
Страница 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Страница 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Страница 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Страница 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Страница 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...
Страница 202: ...Table of contents A EMC declaration APPENDIX 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 2 ...