Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The VICTUS™ Femtosecond Laser Platform is intended for use in the electro‐
magnetic environment specified below. The customer or user of the
VICTUS™ Femtosecond Laser Platform should assure that it is used in such
an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
6 kV contact
±
8 kV air
±
6 kV contact
±
8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast transient /
burst IEC 61000-4-4
±
2 kV for power supply
lines
±
1 kV for input /
output lines
±
2 kV for power supply
lines
±
1 kV for input /
output lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surges IEC 61000-4-5
±
1 kV differential mode
±
2 kV common mode
±
1 kV differential mode
±
2 kV common mode
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
<5 % U
T
(>95 % dip in
U
T
) for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in
U
T
) for 5 sec
<5 % U
T
(>95 % dip in
U
T
) for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in
U
T
) for 5 sec
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
VICTUS™ Femtosecond Laser Plat‐
form requires continued operation
during power mains interruptions, it
is recommended that the VICTUS™
Femtosecond Laser Platform is pow‐
ered from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz) Magnetic
field IEC 61000-4-8
Magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commer‐
cial or hospital environment.
Note: U
T
is the AC mains voltage prior to the application of the test level.
APPENDIX
13
SKU: 70005780TPV * Version 6 * UM-100010770 GB
13 - 4
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 204 of 207
Содержание Victus
Страница 1: ...User Manual VICTUS Femtosecond Laser Platform SW V3 3 SP02 UM 100010770 GB SKU 70005780TPV Version 6 ...
Страница 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Страница 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Страница 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Страница 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Страница 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...
Страница 202: ...Table of contents A EMC declaration APPENDIX 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 2 ...