8.5.5
Parameters for flap
Precondition:
n
A flap procedure must be selected (Fig. 74).
For general information about how to view or specify parameter values refer to
Ä
Chapter 8.5.1 ‘Parameters for capsulotomy’ on page 8 - 36. Parameter-specific
information are described below.
Fig. 74: Dialog boxes for flap procedures
Parameter dialog box
Definition
‘Flap thickness [µm]:’
Distance from the bed to the epithelium sur‐
face measured in µm.
‘Diameter [mm]:’
The diameter of the flap is determined by the
cutting circle of the bed and the rim (meas‐
ured in chord length).
Parameter dialog box
Definition
‘Energy [µJ]:’
Energy level of each single pulse when cutting
the bed.
‘Spot spacing [µm]:’
Distance from spot to spot on the line.
‘Line spacing [µm]:’
Distance of the spots from line to line.
‘Disable bed cut’
To disable the bed cut, click
‘Disable bed cut’
.
‘Flap size’ dialog box
Fig. 75:
‘Flap size’
dialog box
‘Bed’ dialog box
Fig. 76:
‘Bed’
dialog box
SOFTWARE
8
SKU: 70005780TPV * Version 6 * UM-100010770 GB
8 - 48
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 114 of 207
Содержание Victus
Страница 1: ...User Manual VICTUS Femtosecond Laser Platform SW V3 3 SP02 UM 100010770 GB SKU 70005780TPV Version 6 ...
Страница 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Страница 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Страница 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Страница 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Страница 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...
Страница 202: ...Table of contents A EMC declaration APPENDIX 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 2 ...