3.
If the pupil margin has not been detected correctly or if you want to center
the treatment either to the limbus or to the lens apex, proceed as follows:
The following step is mandatory, because the pupil margin defines the
limit for the laser cuts. Please be aware that the laser cannot be focussed
behind the iris!
n
To determine the pupil margin, click the
‘Mark pupil’
button (Fig. 136)
and mark three points on the pupil in the video camera image.
The following step is optional. In case the limbus centering is different to
the pupil centering, the capsulotomy center will be shifted towards the
center defined by the limbus marking. In any case the pupil margin
including the minimum safety distance of 500 µm will be respected.
n
To determine the limbus margin, click the
‘Mark limbus’
button
(Fig. 136) and mark three points on the limbus in the video camera
image.
n
To center to the lens apex, click the
‘Center to lens apex’
button
(Fig. 136) in the video camera image.
Fig. 137: Pupil margin is determined (example)
ð
The green marking circle appears in the video camera image (Fig. 137).
The pupil margin is determined.
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 33
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 159 of 207
Содержание Victus
Страница 1: ...User Manual VICTUS Femtosecond Laser Platform SW V3 3 SP02 UM 100010770 GB SKU 70005780TPV Version 6 ...
Страница 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Страница 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Страница 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Страница 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Страница 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...
Страница 202: ...Table of contents A EMC declaration APPENDIX 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 2 ...