NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, and electromagnetic site
survey should be considered. If the measured field strength in the location in which the
VICTUS™ Femtosecond Laser Platform is used exceeds the applicable RF compliance level above, the
VICTUS™ Femtosecond Laser Platform should be monitored to verify normal operation. If abnormal performance
is observed, additional measures may be necessary, for instance re-orienting or relocating the
VICTUS™ Femtosecond Laser Platform.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
4.
Recommended Separation Distances between Portable and Mobile RF
Communications Equipment and the
VICTUS™ Femtosecond Laser Platform
The VICTUS™ Femtosecond Laser Platform is intended for use in an electro‐
magnetic environment in which radiated RF disturbances are controlled.
The customer or user of the VICTUS™ Femtosecond Laser Platform can help
prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmit‐
ters) and the VICTUS™ Femtosecond Laser Platform as recommended
below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of transmitter (W)
Separation distance according to frequency of transmitter meters (m)
150 kHz to 80 MHz
d = 1.2
Ö
P
80 MHz to 800 MHz
d = 1.2
Ö
P
800 MHz to 2.5 GHz
d = 2.3
Ö
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
APPENDIX
13
SKU: 70005780TPV * Version 6 * UM-100010770 GB
13 - 6
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 206 of 207
Содержание Victus
Страница 1: ...User Manual VICTUS Femtosecond Laser Platform SW V3 3 SP02 UM 100010770 GB SKU 70005780TPV Version 6 ...
Страница 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Страница 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Страница 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Страница 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Страница 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...
Страница 202: ...Table of contents A EMC declaration APPENDIX 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 2 ...