6.2.3
Start-up panel
Fig. 9: Start-up panel
Pos. Function
1
USB ports
2
System key-operated ‘Off’ / ‘On’ switch
3
Indicator for the power supply
4
‘Laser Emission’ indicator during a running procedure.
5
‘Reset’ button to unlock the laser system if the ‘Laser Stop’ button has been pressed, or after a system
error and also by starting the laser system.
6
‘Laser Stop’ button with integrated ‘Bed Stop’ function.
Property damage due to use of unauthorized accessories / tools / devices!
Only USB memory sticks specified in this user manual must be inserted into
the USB ports.
– Do not insert other tools than the one specified in this user manual into
the USB ports. Failure to observe these instructions may damage the
laser system.
6.2.3.1
Switching the laser system on and off
1.
Turn the system key (Fig. 9/2) to the ‘On’ position. The ‘Power OK’ button
(Fig. 9/3) lights up.
ð
A system self test runs for 5-10 minutes.
NOTICE
Switching on
HARDWARE
6
SKU: 70005780TPV * Version 6 * UM-100010770 GB
6 - 4
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 52 of 207
Содержание Victus
Страница 1: ...User Manual VICTUS Femtosecond Laser Platform SW V3 3 SP02 UM 100010770 GB SKU 70005780TPV Version 6 ...
Страница 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Страница 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Страница 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Страница 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Страница 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...
Страница 202: ...Table of contents A EMC declaration APPENDIX 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 2 ...