Treatment
Danger of injury due to user negligence!
The user is responsible for performing a treatment with the VICTUS™ Femto‐
second Laser Platform. The user must supervise the whole treatment
process.
Only the user (surgeon) is allowed to start a treatment by pressing the
‘Start’
button in the GUI and to operate the ‘Procedure’ footswitch. Failure to
observe this responsibility may result in serious permanent patient injury.
Patients with pacemakers
The cover of the laser is designed in a way that electromagnetic fields are minimized.
Nevertheless, special precaution must be taken when treating patients with pace‐
makers. Patients with pacemakers should only be treated in the presence of an anes‐
thetist and after consultation with their cardiologist.
In the case of emergency shutdown of the laser system or treatment abortion by the
user, please refer to
Ä
Chapter 10 ‘Troubleshooting chart’ on page 10 - 1.
9.1
Preparing the laser system for a treatment day
1.
Disinfect the surgery environment.
2.
Switch on the laser system (
Ä
Chapter 6.2.3.1 ‘Switching the laser system on
and off’ on page 6 - 4
).
While the laser system warms up, you can proceed with the following tasks.
3.
Reference the patient bed (
Ä
Chapter 6.2.11.1 ‘Referencing the patient bed’
on page 6 - 12
).
4.
Adjust the surgical microscope, if required (
Ä
Chapter 6.2.9.1 ‘Adjusting the
surgical microscope illumination’ on page 6 - 10
).
5.
When the laser system warm-up is finished, login to the software.
6.
Select the treatment mode that you want to perform (Fig. 27/4).
7.
Perform the system tests (
Ä
Chapter 9.3 ‘Performing the daily system test’
on page 9 - 2
).
9.2
Maintaining the laser system after a treatment
day
1.
Clean the spacer cone (
Ä
‘Cleaning the spacer cone’ on page 6 - 14
).
2.
Protect the spacer cone with the protection cover (
Ä
Chapter 6.2.8.1 ‘Pro‐
tecting the spacer cone’ on page 6 - 8
).
3.
Clean the patient bed (
Ä
‘Cleaning the patient bed’ on page 6 - 14
).
L
CAUTION
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 1
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 127 of 207
Содержание Victus
Страница 1: ...User Manual VICTUS Femtosecond Laser Platform SW V3 3 SP02 UM 100010770 GB SKU 70005780TPV Version 6 ...
Страница 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Страница 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Страница 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Страница 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Страница 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...
Страница 202: ...Table of contents A EMC declaration APPENDIX 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 2 ...