25.
Click the
‘Continue’
button (Fig. 150) to open the ‘Treatment result’ screen
that shows the procedure results and the number of remaining licenses.
Fig. 151:
‘LRCS’
‘Treatment result’
screen
ð
L
CAUTION! Danger of trauma due to lowering the patient bed
without having released the vacuum of the suction clip!
When the vacuum is activated, lowering the patient bed leads to
an abrupt loss of vacuum and causes pain to the patient.
– Do not lower the patient bed while the vacuum is activated and
the patient is docked.
26.
Do not remove the suction clip while the vacuum is activated.
27.
Verify that the eye suction has been released after the procedure has been
executed.
ð
The ‘Eye suction’ status indicator of the suction clip turns gray
.
If a treatment is completed and the
‘Eye suction’
status indicator does not
turn gray, you must manually pull off the suction reservoir from the suction
port at the laser system.
28.
Disable the ‘PI suction’ in the software to deactivate the vacuum of the
spacer cone.
29.
Complete the cataract treatment (
Ä
Chapter 9.4.6 ‘Completing a treatment’
on page 9 - 29
).
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 43
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 169 of 207
Содержание Victus
Страница 1: ...User Manual VICTUS Femtosecond Laser Platform SW V3 3 SP02 UM 100010770 GB SKU 70005780TPV Version 6 ...
Страница 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Страница 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Страница 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Страница 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Страница 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...
Страница 202: ...Table of contents A EMC declaration APPENDIX 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 2 ...