Danger of confusion due to identical-looking connectors at the interlock
cable!
– Check for the labels on the cable to identify the connector that corre‐
sponds to the respective interface port.
1.
Connect the laser system and the patient bed. Observe the cable labeling.
2.
Verify that all cables are connected correctly and are seated firmly. The
power cable must be snapped into place.
3.
Verify the proper function of the interlock connection. To do so, apply a ver‐
tical force of
>
6 N. Observe the LED indicators on the force display. The
patient bed must only be movable in the downward direction.
4.2
Service and calibration
Hazard due to improperly completed service or calibration!
Service or calibration that is completed by personnel other than Authorized
Service Technicians may result in severe injuries to persons who perform the
service, third parties or patients, and may damage the laser system.
– Service and calibration must only be performed by Authorized Service
Technicians.
If system checks fail or other errors occur, service must be performed as soon as pos‐
sible.
The service and calibration interval is 6 months.
The expected service life is 5 years.
4.3
Decomissioning and disposal
Danger of injury in conjunction with improper decommissioning and dis‐
posal!
Stored residual energies and components featuring edges, tips, and corners
on and inside the laser system or on the required tools may cause injuries.
– Decommissioning and disposal must be performed by Authorized
Service Technicians exclusively.
Technolas Perfect Vision GmbH will arrange for the collection and recycling of (a)
end-of-life Technolas Perfect Vision GmbH equipment and (b) end-of-life equip‐
ment from other manufacturers when a Technolas Perfect Vision GmbH product
is purchased to replace the older equipment and has identical functionality. To
initiate the collection of end-of-life equipment, contact an Authorized Service
Technician.
NOTICE
L
WARNING
When do perform service
and calibration
Service life
L
WARNING
INSTALLATION, COMMISSIONING, SERVICE, CALIBRATION, DECOMISSIONING AND
DISPOSAL
4
SKU: 70005780TPV * Version 6 * UM-100010770 GB
4 - 3
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 41 of 207
Содержание Victus
Страница 1: ...User Manual VICTUS Femtosecond Laser Platform SW V3 3 SP02 UM 100010770 GB SKU 70005780TPV Version 6 ...
Страница 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Страница 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Страница 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Страница 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Страница 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...
Страница 202: ...Table of contents A EMC declaration APPENDIX 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 2 ...