2.10.1
Labels in the operating room entrance area
The following labels must be displayed in the operating room entrance area:
Label name
Label
Description
No smoking
Fire hazards exist from flammable or potentially explo‐
sive solid, liquid, or gaseous substances.
Unauthorized
access prohibited
Only persons authorized by the customer may enter the
operating room.
Laser beam
Unauthorized persons must not enter work places
where laser beams are used and they are not permitted
to operate laser emitting equipment.
The equipment should only be operated when the user
is sure that the laser output cannot damage eyes, burn
skin or tissue, or result in damage to protective work
clothing or other objects.
Never look directly into the laser source!
Invisible laser radia‐
tion (large)
INVISIBLE LASER RADIATION
AVOID EXPOSURE TO BEAM
LASER CLASS 3B
Wavelength: 1040
±
25 nm
Pulse duration: 290
–
550 fs
Maximum pulse frequency: 160 kHz
Maximum output power: 0.86 W
IEC 60825-1:2014
SAFETY
2
SKU: 70005780TPV * Version 6 * UM-100010770 GB
2 - 12
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 28 of 207
Содержание Victus
Страница 1: ...User Manual VICTUS Femtosecond Laser Platform SW V3 3 SP02 UM 100010770 GB SKU 70005780TPV Version 6 ...
Страница 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Страница 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Страница 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Страница 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Страница 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...
Страница 202: ...Table of contents A EMC declaration APPENDIX 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 2 ...