9.
Verify that the eye is correctly centered.
10.
Verify that one of the (more central) green or yellow shear force LEDs on
the force display is illuminated.
n
If not, correct the pressure by using the X/Y-joystick (Fig. 134/1) for the
shear force until one LED is illuminated green or yellow.
11.
Close the suction clip by pressing the ends of the suction clip together.
ð
The patient is docked to the PI 125.
12.
If the pressure status display does not turn green, move the patient bed
until you reach the green pressure status display area. Ensure that there are
no wrinkles visible in the OCT image.
L
CAUTION!
Ensure that the pressure status display does not reach the upper red
pressure status display area! Otherwise there is a higher risk to have
posterior wrinkles.
13.
Verify again that one of the (more central) green or yellow shear force LEDs
on the force display is still illuminated.
14.
If the docking was performed correctly, a thin fluid film is visible in the
periphery between the cornea and the PI 125, and the cornea preserves its
natural shape.
9.4.4.3
For all other procedures using a headrest with integrated joy‐
stick
1.
Move the patient bed using the joystick (Fig. 132/1 or Fig. 133/1) of the
patient bed control at the headrest until the suction clip and the PI 125 are
aligned.
2.
Slowly move the patient bed upward until the suction clip almost touches
the PI 125.
3.
Be aware that the reflection of the laser ring light is visible in the video camera
image.
Ensure that the laser ring light illumination switch is adjusted to position
‘10’
(Fig. 12/1). Align the reflection of the laser ring light on the cornea with
the laser ring light and gently move the patient bed upward.
4.
If the pressure status display does not show green, move the patient bed
until you reach the green pressure display area.
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 26
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 152 of 207
Содержание Victus
Страница 1: ...User Manual VICTUS Femtosecond Laser Platform SW V3 3 SP02 UM 100010770 GB SKU 70005780TPV Version 6 ...
Страница 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Страница 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Страница 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Страница 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Страница 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...
Страница 202: ...Table of contents A EMC declaration APPENDIX 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 2 ...