4.
Slowly move the patient bed upward until the PI 125 aligns completely
with the corneal surface.
ð
The shear force display shows the lateral pressure that is exerted on the
eye. The optimal position is achieved when the LEDs are illuminated
green.
L
WARNING! Danger of injury due to contaminations!
There must be no air bubbles, excess fluids, etc. between the PI 125 and the
corneal surface. This situation may impair the treatment and cause severe
patient injuries.
5.
Check if there are air bubbles, fluids, or impurities between the PI 125 and
the cornea.
It may be useful to change the camera focus between the lens and the cornea
to better recognize air bubbles.
n
If you see air bubbles, fluids, or impurities between the PI 125 and the
cornea, undock the patient.
n
If the PI 125 is clean, continue with step 8.
6.
Repeat steps 1 to 5 until no air bubbles, fluids, or impurities are visible
between the PI 125 and the cornea.
7.
Verify that there is no water ring between the PI 125 and the cornea.
n
If there is a water ring, undock the patient and use a Merocel
®
sponge
to absorb excess fluids. Repeat steps 1 to 7.
8.
Verify that the eye is correctly centered.
9.
Verify that one of the (more central) green or yellow shear force LEDs on
the force display is illuminated.
n
If not, correct the pressure by using the X/Y-joystick for the shear force
and the Z-joystick for the vertical force until one LED is illuminated
green or yellow.
10.
Close the suction clip by pressing the ends of the suction clip together.
ð
The patient is docked to the PI 125.
11.
Verify again that one of the (more central) green or yellow shear force LEDs
on the force display is still illuminated.
9.4.5
Repositioning the suction clip
If you have to reposition the suction clip, proceed as follows:
1.
Open the suction clip by pulling the snap-fit backwards.
2.
Move the patient bed slowly downward until there is enough space to
reposition the suction clip without touching the PI 125 or the spacer cone.
3.
Release the vacuum of the suction clip by pressing the ‘Vacuum’ foot‐
switch once.
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 28
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 154 of 207
Содержание Victus
Страница 1: ...User Manual VICTUS Femtosecond Laser Platform SW V3 3 SP02 UM 100010770 GB SKU 70005780TPV Version 6 ...
Страница 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Страница 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Страница 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Страница 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Страница 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...
Страница 202: ...Table of contents A EMC declaration APPENDIX 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 2 ...