12.
If you want to accept all procedure workflow settings, click the
‘Confirm’
button (Fig. 142).
The user must verify that the parameter setting has been performed correctly.
Fig. 143: Pattern alignment for the cataract treatment accepted
ð
The color marking of the circle in the video camera image and the out‐
lines in the linear scan images turn green. The
‘Start’
button becomes
available (Fig. 143).
L
WARNING! Risk of tissue bridges due to movement when laser is
firing!
Tissue bridges may occur if vibrations from movement occur when
the laser is firing.
– Do not walk around while the procedure is performed.
– Do not touch the patient bed or the patient's head while the
procedure is performed.
– Do not touch the suction clip.
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 37
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 163 of 207