18.
To change the preset pachymetry value, click the
‘Edit’
button in the ‘AK @
225°’ OCT B scan image (Fig. 147).
Fig. 148:
‘Arcuate incision editor’
(example)
19.
L
CAUTION!
Be aware that the pachymetry values can vary over the length of the
arcuate incisions cut.
To change the pachymetry value to the patient's actual pachymetry, left-
click on the yellow dot (Fig. 148) and drag and drop it to the rear side of the
cornea.
Fig. 149: Changing the pachymetry value
ð
The ‘Pachymetry [µm]:’ value changes. Simultaneously, the ‘Resulting
depth [µm]:’ value changes based on the preset ‘Depth ratio [%]:’ per‐
centage.
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 41
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 167 of 207