Infection hazard due to loss of sterility!
The PI 125 Kit packaging must not be damaged or opened. The expiry dates
must not be exceeded. Damaged or opened packaging may result in loss of
sterility and increases the risk of infections and severe injuries.
– Always inspect the PI 125 Kit packaging and ensure that it is not dam‐
aged in any way.
– Do not use the PI 125 and the suction clip if packaging is damaged or
already opened.
– Do not open the packaging until you are ready to use it.
– Do not touch the PI 125 and the suction clip with powdered medical
gloves. Only open shortly before treatment.
Infection hazard due to loss of sterility!
The non-sterile person must not touch the PI 125 or the suction clip. If neces‐
sary, dispose of the PI 125 Kit and use a different one.
The sterile person must not touch the packaging of the PI 125 Kit.
Before connecting the PI 125 to the spacer cone, ensure that the contact sur‐
face of the spacer cone is absolutely clean.
– Clean the contact surface of the spacer cone, if necessary (
Ä
‘Cleaning
the spacer cone’ on page 6 - 14
).
1.
Inspect the PI 125 Kit packaging, and dispose of all materials if they are
damaged or already open.
2.
Follow the procedures below.
9.4.3.1
Connecting the PI 125 (REF 90000115) to the spacer cone and
positioning the suction clip (REF 90000115)
WARNING! Ensure that the PI 125 (REF 90000115) is only used together with the
suction clip (REF 90000115)!
WARNING! Ensure that the suction clip (REF 90000115) is only used together
with the PI 125 (REF 90000115)!
L
WARNING
L
WARNING
L
CAUTION
TREATMENT
9
SKU: 70005780TPV * Version 6 * UM-100010770 GB
9 - 13
TPV Controlled Document / Status: Effective / Effective Date: Jun 14, 2017 2:38:46 PM
UM-100010770 / Version 6 / Page 139 of 207
Содержание Victus
Страница 1: ...User Manual VICTUS Femtosecond Laser Platform SW V3 3 SP02 UM 100010770 GB SKU 70005780TPV Version 6 ...
Страница 12: ...GENERAL INFORMATION A SKU 70005780TPV Version 6 UM 100010770 GB A 4 ...
Страница 16: ...INTRODUCTION 1 SKU 70005780TPV Version 6 UM 100010770 GB 1 4 ...
Страница 48: ...CLINICAL APPLICATIONS 5 SKU 70005780TPV Version 6 UM 100010770 GB 5 6 ...
Страница 186: ...TREATMENT 9 SKU 70005780TPV Version 6 UM 100010770 GB 9 60 ...
Страница 201: ...Appendix 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 1 ...
Страница 202: ...Table of contents A EMC declaration APPENDIX 13 SKU 70005780TPV Version 6 UM 100010770 GB 13 2 ...