Usage:
Query only
Manual operation:
See
CONTrol<Ch>:SEGMent<Nr>:SEQuence:CLEar:ALL
For segmented sweeps this command deletes the command/switch sequence for the
respective channel and sweep segment (defined using
SEGMent<Nr>:SEQuence<Nr>:RFFE<Bus>:COMMand:DATA
and
SEGMent<Nr>:SEQuence<Nr>:GPIO<Port>:VOLTage
).
Suffix:
<Ch>
.
Channel number
<Nr>
Segment number
Usage:
Event
Manual operation:
See
CONTrol<Ch>:SEGMent<Nr>:SEQuence<Nr>:DELay
<Delay>
For segmented sweeps this command allows to introduce delays between subsequent
sequence steps and between the final sequence step and the sweep (segment) start.
The command/switch sequences are defined using
SEQuence<Nr>:RFFE<Bus>:COMMand:DATA
and
SEQuence<Nr>:GPIO<Port>:VOLTage
.
Suffix:
<Ch>
.
Channel number
<Nr>
Segment number
<Nr>
Sequence number, defining the order in which the commands
shall be executed.
For every channel, segment and RFFE interface, the sequence
numbers must be consecutive, starting at 1; reusing a sequence
number overwrites a previously defined command.
Parameters:
<Delay>
Delay time
Default unit: s
Manual operation:
See
"Wait (Sweep Sequencer Table)"
CONTrol<Ch>:SEGMent<Nr>:SEQuence<Nr>:GPIO<Port>:VOLTage
<Voltage>
For segmented sweeps this command allows to define the GPIO voltage settings to be
applied at the start of each segment.
The GPIO ports can be enabled/disabled using
. Complementary RFFE commands can be defined using
SEGMent<Nr>:SEQuence<Nr>:RFFE<Bus>:COMMand:DATA
.