[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault
<TestPort1>, <TestPort2>...
Creates the default port assignment(s) for the specified test ports.
The number of created assignments can be queried using the
]CORRection:COLLect:AUTO:ASSignment:COUNt?
command.
Note
that during the corresponding calibration sweep the R&S
ZNB/ZNBT expects the
physical port connections to be established exactly as specified by the created port
assignments.
Use
●
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:TPORt:
DEFault
on page 1023 to use auto-detection instead of defining "test port --> cal
unit port"-connections explicitely
●
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:
TPORt
to define (additional) port assignments manually
See
Chapter 4.5.5.4, "Multiple Port Assignments"
on page 164 for background infor-
mation.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<TestPort1>
Test port number 1
<TestPort2>
Test port number 2
Example:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:TPORt:DEFault
<TestPort1>, <TestPort2>...
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:
, but initially defines the default port assignments only by their
underlying test port set(s); the connected calibration unit ports are
auto-detected
at
the start of the corresponding calibration sweep.
The number of created assignments can be queried using the
]CORRection:COLLect:AUTO:ASSignment:COUNt?
command.
See
Chapter 4.5.5.4, "Multiple Port Assignments"
on page 164 for background infor-
mation.
Suffix:
<Ch>
.
SCPI Command Reference