Remote Control
R&S
®
ZNB/ZNBT
663
User Manual 1173.9163.02 ─ 38
most significant bit) is set to zero for all parts. Thus the contents of the register parts
can be processed by the controller as positive integer.
The sum bit is obtained from the EVENt and ENABle part for each register. The result
is then entered into a bit of the CONDition part of the higher-order register.
The instrument automatically generates the sum bit for each register. Thus an event
can lead to a
throughout all levels of the hierarchy.
The five parts of an SCPI register have different properties and function as described
below.
CONDition
The CONDition part is permanently overwritten by the hardware or the sum bit of the
next lower register. Its contents always reflect the current instrument state.
This register part can only be read, but not overwritten or cleared. Reading the CONDi-
tion register is nondestructive.
PTRansition
The two transition register parts define which state transition of the condition part
(none, 0 to 1, 1 to 0 or both) is stored in the EVENt part.
The Positive TRansition part acts as a transition filter. When a bit of the CONDition part
is changed from 0 to 1, the associated PTR bit decides whether the EVENt bit is set to
1:
●
PTR bit = 1: the EVENt bit is set
●
PTR bit = 0: the EVENt bit is not set
This status register part can be overwritten and read at will. Reading the PTRansition
register is nondestructive.
Status Reporting System