The load match correction can provide a significant improvement of the transmission
S-parameter measurements if the load ports are poorly matched. With sufficiently
matched load ports, you can disable the correction to gain speed.
Note:
Disable the load match correction if your test setup or DUT is not suited for
reverse sweeps.
Remote command:
[SENSe<Ch>:]FREQuency:CONVersion:GAIN:LMCorrection
Scalar Power Cal – All Power Cals On / All Power Cals Off
Activates or deactivates all scalar power calibrations in the active channel. "All Power
Cals On" is only available if a valid power calibration is available for the active channel,
but not active; see "Channel State" in
Chapter 5.4.4.3, "Calibration Manager Dialog"
Note:
A label "PCal Off" appears behind the trace list of a wave quantitiy or a ratio if
the power calibration is switched off; see also
Chapter 4.5.6.3, "Power Calibration
on page 170. The calibration status of each channel and trace appears in the
setup information (see
Remote command:
Manage Cals – Recall Last Cal Set
Loads and activates the recall set for which the last calibration was performed. If the
last calibrated setup is already active, nothing is changed. The calibrated setups are
automatically stored in the
C:\Users\Public\Documents\Rohde-Schwarz\Vna\Calibration\
RecallSets
directory. A message box pops up if the directory is empty, e.g. because
no calibration was performed yet.
Remote command:
n/a
5.4.4.2
Active Power Cals Dialog
The "Active Power Cals" dialog shows the power calibrations for the active channel,
enables and disables power calibrations.
Access:
CHANNEL – CAL > "Use Cal" > "Active Power Cals..."
Cal Softtool