GUI Reference
R&S
®
ZNB/ZNBT
412
User Manual 1173.9163.02 ─ 38
5.7.2.1
Controls on the Offset Tab
The "Fixture Compensation..." button opens the
Overview
This button is available on all "Offset Embed" softtool tabs. It opens the
.
Port
Physical test port of the analyzer. You can define independent offset parameters for all
ports.
Remote command:
The
<PhyPt>
numeric suffix in the
[SENSe<Ch>:]CORRection:...
commands
identifies the physical port.
Delay / Electrical Length / Mech. Length
Defines the length offset at the selected port as a delay, an electrical length, or a
mechanical length. The three quantities are related by:
r
mech
r
mech
L
Length
Electrical
c
L
Delay
;
and overwrite each other. See also
Chapter 4.6.1.1, "Definition of Offset Parameters "
Note:
The entered parameters must correspond the actual (one-way) length of the
transmission line. To account for the propagation in both directions, the phase shift of a
reflection parameter due to a given length offset is twice the phase shift of a transmis-
Offset Embed Softtool