Getting Started
R&S
®
ZNB/ZNBT
17
User Manual 1173.9163.02 ─ 38
3.1.3 Bench Top Operation
If the analyzer is operated on a bench top, the surface should be flat. The instrument
can be used in horizontal or vertical position, standing on its feet, or with the support
feet on the bottom expanded.
3.1.4 Operation in a 19" Rack
The R&S
ZNB/ZNBT can be mounted in 19" racks using the adapter R&S ZZA-KN5
(order number 1175.3040.00). Proceed according to the mounting instructions supplied
with the rack adapter.
Risk of instrument damage due to insufficient airflow in a rack
If the instrument is run with insufficient airflow for a longer period, the instrument over-
heats, which can disturb the operation and even cause damage.
Make sure that all fan openings are unobstructed, that the airflow perforations are
unimpeded, and that the minimum distance from the wall is 10
cm.
3.1.5 EMI Suppression
Electromagnetic Interference (EMI) may affect the measurement results.
To suppress generated Electromagnetic Interference:
●
Use suitable shielded cables of high quality. For example use double-shielded RF
and LAN cables.
●
Always terminate open cable ends.
●
Note the EMC classification in the data sheet.
3.1.6 Connecting the Analyzer to the AC Supply
The network analyzer is automatically adapted to the AC supply voltage, which must
be in the range of 100 V to 240 V at 50 Hz to 60 Hz. A line frequency of 400 Hz is also
supported. The mains connector is located in the upper part of the rear panel (see
Chapter 3.2.3, "Rear Panel R&S
Putting the Analyzer into Operation