GUI Reference
R&S
®
ZNB/ZNBT
246
User Manual 1173.9163.02 ─ 38
The measurement is triggered according to the current trigger settings (see
on page 400). Each trigger event triggers the first partial mea-
surement of a measurement point. The time interval between two consecutive mea-
surements depends on the trigger settings and the sweep parameters (especially the
number of points). Any trigger mode is allowed.
A "CW Mode" sweep corresponds to the analysis of a signal over the time with a time
scale and resolution that is determined by the trigger events. In a Cartesian diagram
the measurement result is displayed as a trace over a linear time scale (like e.g. in an
oscilloscope). The diagram is similar to the "Time" diagram. The following example
shows a "CW Mode" sweep with a DUT that does not markedly change its transmis-
sion characteristics over the time.
Tip: Sweep time
The time interval between two consecutive trigger pulses must not be smaller than the
minimum measurement time per measurement point. See
Remote command:
Time
sweeps, are performed at constant frequency and stim-
ulus power, which can be specified in the "Stimulus" settings (see
on page 430). A single sweep extends over a specified period of time,
defined via the "Stop Time" setting. The time intervals between two consecutive sweep
points are calculated according to "Stop Time"/(n - 1) where n is the selected
.
A "Time" sweep corresponds to the analysis of a signal over the time; the function of
the analyzer is analogous to an oscilloscope. In a Cartesian diagram the measurement
result is displayed as a trace over a linear time scale. The following example shows a
"Time" sweep with a DUT that does not markedly change its transmission characteris-
tics over the time.