Command Reference
R&S
®
ZNB/ZNBT
984
User Manual 1173.9163.02 ─ 38
Enable the MCTYpes mode before defining the calibrations to be performed (using
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
, disable it after the calibra-
tions were saved (using
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected[:
once
). Also make sure to call
SENSe<Ch>:CORRection:COLLect:SAVE:SELected
with
<Ch>
pointing to one of
the calibrated channels.
Parameters:
<Boolean>
Example:
Suppose there are 3 channels in the current recall set and you
want to calibrate channels 1 and 2. For channel 1 you want to
perform a Reflection Normalization (Short) on ports 1 and 2, for
channel 2 an OSM calibration on port 2. This requires measuring
standards as shown in the table below. Proceed as follows:
:CORRection:COLLect:CHANnels:MCTYpes ON
Activate the MCTYPes mode.
:SENSe1:CORRection:COLLect:METHod:DEFine
'Parser Test SFK Ch1', RSHort, 1, 2
:SENSe2:CORRection:COLLect:METHod:DEFine
'Parser Test SFK Ch2', FOPort, 2
Declare the calibrations to be performed, then connect the Short
standard to port 1
:SENSe1:CORRection:COLLect:ACQuire:SELected
SHORt, 1
Acquire calibration data for channel 1, then connect the Open
standard to port 2
:SENSe2:CORRection:COLLect:ACQuire:SELected
OPEN, 2
Acquire calibration data for channel 2, then connect the Short
standard to port 2
:SENSe1:CORRection:COLLect:ACQuire:SELected
SHORt, 2
:SENSe2:CORRection:COLLect:ACQuire:SELected
SHORt, 2
Acquire calibration data for channel 1 and 2, then connect the
Match standard to port 2
:SENSe2:CORRection:COLLect:ACQuire:SELected
MATCh, 2
Acquire calibration data for channel 2
:SENSe1:CORRection:COLLect:SAVE:SELected
Complete all calibrations. This command is required only once
and has to address one of the calibrated channels (channel 1 or
2 in this example)
:CORRection:COLLect:CHANnels:MCTYpes OFF
Deactivate the MCTYPes mode. Recommended for compatibility
with manual operation!
SCPI Command Reference