GUI Reference
R&S
®
ZNB/ZNBT
310
User Manual 1173.9163.02 ─ 38
Messages in the dialog
An information (or error message) is displayed if one of the following happens:
●
One of the selected calibration kits is described by ideal kit parameters or typical
values.
●
One of the selected calibration kits does not contain all standards that are required
for the previously selected calibration type.
●
Different connector types are defined at the ports but the selected calibration type
requires uniform connectors.
●
A cal kit standard does not cover the entire calibrated frequency range.
●
The selected power meter does not cover the entire calibrated frequency range.
The layout of the main panel depends on the number of test ports available.
Figure 5-21: SMARTerCal(Manual), Step 2: Connectors and Cal Kits (>4 ports)
The upper part of the panel shows the previously configured port(s) and calibration
type. The lower part gives access to the connector and cal kit settings.
Connector / Gender
Defines the connector types and genders of the ports to be calibrated. For symmetric
(sexless) connector types (e.g. 7 mm / PC7), "Gender" is unavailable.
If "Same Connector All Ports" is active, the connector types at all ports (but not their
gender) are always adjusted to the current selection. If "Same Gender All Ports" is
active, the genders at all ports are always adjusted to the current selection.
User-defined connectors can be added or removed in the
.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:CONNection<PhyPt>
[SENSe<Ch>:]CORRection:COLLect:SCONnection<PhyPt>
[SENSe<Ch>:]CORRection:CONNection
Cal Softtool