GUI Reference
R&S
®
ZNB/ZNBT
508
User Manual 1173.9163.02 ─ 38
Imag
Real
10
180
/
Phase
dB
20
/
Magnitude
j
e
a
M
M
j
old
new
The formula and the different constants are adjusted to the different display formats of
a trace:
●
The "Mag" (nitude) factor shifts a dB Mag trace in vertical direction, leaving the
phase of a complex parameter unchanged.
●
The "Phase" factor rotates a trace that is displayed in a polar diagram around the
origin, leaving the magnitude unchanged.
●
The "Real" value shifts a real trace in vertical direction, leaving the imaginary part
unchanged.
●
The "Imag" (inary) value added constant shifts an imaginary trace in vertical direc-
tion, leaving the real part unchanged.
Tip:
Shifting the trace by means of constant values is a simple case of trace mathe-
matics. Use the "User Def Math" dialog to define more complicated mathematical oper-
ations (see
Chapter 5.12.2.2, "User Def Math Dialog"
Remote command:
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y:OFFSet
5.12.8 Trace Data Tab
Stores one or several data or memory traces to a file or loads a memory trace from a
file.
Background information
Refer to
Trace Config Softtool