GUI Reference
R&S
®
ZNB/ZNBT
291
User Manual 1173.9163.02 ─ 38
Start Cal Sweep
Starts the calibration sweep(s) for the selected port and power calibration settings and
closes the dialog. The calibration is performed as described in
Open the
if you wish to modify the calibration procedure.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect[:ACQuire]
Power Cal Dialog (Source Flatness)
In "Source Flatness" mode the "Power Cal" dialog guides you through the source
power calibration for a single port. The source flatness data complement an active
"SMARTerCal".
Access:
CHANNEL – CAL > "Start Cal" > "Power Cal..." > "Source Flatness"
Background information
Refer to
Chapter 4.5.6.1, "Source Power Calibration"
ter 4.5.7.3, "Combining SMARTerCal with Scalar Power Calibration"
Port Overview
The dialog shows the selected source port with its current cal power settings (see
Chapter 5.4.3.2, "Modify Cal Power Dialog"
on page 334); moreover, a circuit diagram
visualizes the purpose of the flatness and reference receiver calibration. The DUT or a
Match must be connected to the calibrated port.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Cal Softtool