ELECTROMAGNETIC TYPE

DIGITAL COATING THICKNESS METER

SM‑1100

INSTRUCTION

MANUAL

CAUTIONS:

●Before using the Meter, read this INSTRUCTION MANUAL thoroughly

and use the Meter correctly.

●Keep this INSTRUCTION MANUAL carefully and refer to this, when

necessary.

●In the event of any doubt arising, the original INSTRUCTION MANUAL

in Japanese is to be final authority.

※Sensitivity of the probe of this Meter has been adjusted so as to

deal with the measurement of a thicker film thickness.

When the probe is left on a metal e.g. steel desk etc. the Meter may
indicate a certain value or「HHH」occasionally.

SANKO ELECTRONIC LABORATORY CO., LTD.

(株式会社 サンコウ電子研究所)

Summary of Contents for SM-1100

Page 1: ...efer to this when necessary In the event of any doubt arising the original INSTRUCTION MANUAL in Japanese is to be final authority Sensitivity of the probe of this Meter has been adjusted so as to dea...

Page 2: ...7 4 1 Starting of the statistical function and alteration of 15 the indicating data 7 4 2 Statistical measurement mode 15 7 4 3 Indication of the measured numbers 16 7 4 4 Indication of the mean valu...

Page 3: ...nt of the thickness of a non conductive film such as coating lining etc applied to iron substrate including steel and ferritic stainless steel SUS 430 etc The Meter can be applicable for various domes...

Page 4: ...nal functions 1 Conversion of the key lock mode 2 Setting of the upper lower limit values 3 Memory of 1800 measurement values 4 Statistical calculating functions such as measurement numbers mean value...

Page 5: ...l measurement mode Also for converting the indicating kinds of statistical data in the statistical measurement mode DATA OUT key Key for outputting the stored data to the connected printer in the stat...

Page 6: ...move rust dirt or dust if any 5 2 Connecting Disconnecting the probe Carry out the connection disconnection of the probe by the following procedures Be sure to keep the Power OFF The probe cannot be s...

Page 7: ...The buzzer emits a beeping sound The Meter is ready for measurement When power ON the last measured value of the previous measurement is indicated When the first power ON or all data are erased in the...

Page 8: ...d on the LCD the adjustment is acceptable When the indications are away from 0 repeat above procedures several times When LLL is indicated while zero adjusting its adjusting point is widely deviated f...

Page 9: ...s possible for both the probe is contacting to the standard thickness plate or not When the key is pressed CAL is indicated on the upper left of the LCD with a beeping sound of the buzzer and indicati...

Page 10: ...to the next item 6 5 Resetting the working curve When the new values are set for both zero adjustment and standard calibration the last characteristic of substrate working curve is erased and new set...

Page 11: ...e LCD and 0 is indicated blinking The buzzer emits a beeping sound for every press of the key The buzzer emits beeping sounds E is indicated for 3 seconds The Meter is restored to the possible conditi...

Page 12: ...f the resuming function at the succeeding power ON Correct usage Incorrect usage Hold the mid part of the probe and Don t tilt the probe Don t slide the probe Don t press the probe press it down verti...

Page 13: ...the key lock mode Press the LOCK key in power ON condition Then the buzzer emits 3 beeping sounds All keys except the power key can not be operated to prevent from mis operation To release the key loc...

Page 14: ...it value is 1 80mm is not changed in case 2 1 80mm is indicated blinking Press the H L key The buzzer emits a beeping sound L is indicated blinking for about 3 seconds Indication changes after about 3...

Page 15: ...r the measurement Press the H L key The buzzer emits a beeping sound Erasing the upper and lower limit values are cancelled No set upper and lower limit values are erased The Meter is restored to the...

Page 16: ...sound and the mode proceeds to the normal measurement mode 7 4 2 Statistical measurement mode Measured data up to maximum 1800 points can be stored in memory in the measurement order and Measured num...

Page 17: ...ue Press the STAT key 3rd time from the normal measurement mode The buzzer emits a beeping sound E and the mean value XXX are indicated alternately Indicating time of E about 0 5 seconds Indicating ti...

Page 18: ...nately Indicating time of L about 0 5 seconds Indicating time of XXX about 2 seconds 7 4 7 Indication of the standard deviation value Press the STAT key 6th time from the normal measurement mode The b...

Page 19: ...No When key is pressed the indication Memory No traverse quickly by keeping of the data move down from the last memory No on key pressing When key is pressed again after suspending the key operation N...

Page 20: ...hile indicating on the LCD Data can be erased even while indicating the memorized data on the LCD Refer to the later item 10 2 Erasing the last data together Memory No NNN and data XXX are indicated a...

Page 21: ...f the memorized data EEE is indicated In case of cancelling of the erasing all memorized data After erasing the mode returns to the possible condition for the memory measurement in the initial conditi...

Page 22: ...returns to the possible condition for the memory measurement Unit m or mm is indicated blinking The data just before erased data can be erased one after another by each press of the DATA CLR key Press...

Page 23: ...batteries Battery Dry battery R03 1 5V 4 pcs MAINTENANCE AND INSPECTION Use the Meter within the range of 0 to 40 Avoid to expose the Meter to dew condensation wet with water dust intense heat vibrati...

Page 24: ...can not be measured occasionally With respect to an elastic film place a standard thickness plate with about 0 5 mm on it and subtract the thickness of the standard thickness plate from the total thi...

Page 25: ...ector Iron piece detector Manufacturer Sanko Electronic Laboratory Co Ltd 2 6 4 Tokyo branch Shibata Bldg Uchikanda Chiyoda ku Tokyo Japan 101 0047 81 3 3254 5031 81 3 3254 5038 Tel Fax Osaka branch K...