GUI Reference
R&S
®
ZNB/ZNBT
332
User Manual 1173.9163.02 ─ 38
5.4.3 Pwr Cal Settings Tab
Provides access to all functions for power meter and power calibration data handling
(transmission coefficients). Power calibration of the R&S
ZNB/ZNBT is a fully menu-
guided process.
Efficient power calibration procedure
●
For standard applications, open the "Start Cal" tab and select "Scalar Power Cal" –
"Power Cal..." to perform the necessary calibration sweeps with default power cali-
bration settings. You do not need any of the buttons in the "Pwr Cal Settings" tab.
●
Select "Cal Power..." if you use an amplifier between the source port and the DUT.
●
Select "Transm. Coefficients..." if you want to modify the power calibration proce-
dure.
Background information
Refer to
Chapter 4.5.6, "Scalar Power Calibration"
Some buttons serve as openers for additional dialogs:
●
"Cal Power..."
: see
Chapter 5.4.3.2, "Modify Cal Power Dialog"
●
"Transm. Coefficients..."
Chapter 5.4.3.3, "Power Meter Transmission Coef-
●
"Power Meters...": see
Chapter 4.7.18, "External Power Meters"
5.4.3.1
Controls on the Pwr Cal Settings Tab
Cal Softtool