CALCulate<Chn>:LIMit:SEGMent<Seg>:AMPLitude:STARt
<Response>
CALCulate<Chn>:LIMit:SEGMent<Seg>:AMPLitude:STOP
<Response>
These commands change the start or the stop response values (i.e. the response val-
ues assigned to the start or stop stimulus values) of a limit line segment. A segment
must be created first to enable the commands (e.g
See also
Chapter 4.4.1.1, "Rules for Limit Line Definition"
Tip:
To define the response values of several limit line segments with a single com-
CALCulate<Chn>:LIMit:LOWer[:DATA]
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg>
Segment number
Parameters:
<Response>
Response value
Range:
Virtually no restriction for limit segments
*RST:
The default response values of a segment that is
created by defining its stimulus values only (e.g. by
means of CALCulate<Ch>:LIMit:CONTrol[:DATA]),
are -40 dB.
Default unit: NN
Example:
CALC:LIM:DATA 1,1500000000, 2000000000,2,3
Define an upper limit line segment (segment no. 1) in the stimu-
lus range between 1.5 GHz and 2 GHz, assigning response val-
ues of +2 dB and +3 dB.
:CALC:LIM:SEGM:AMPL:STAR 5; STOP 5; :CALC:LIM:
SEGM:TYPE LMIN
Change the segment to a lower limit line segment with a con-
stant response value of +5 dB.
CALC:LIM:DATA?
Query the type, the stimulus and response values of the created
segment with a single command. The response is
2,1500000000,2000000000,5,5.
Manual operation:
See
CALCulate<Chn>:LIMit:SEGMent<Seg>:STIMulus:STARt
<FreqPowTime>
CALCulate<Chn>:LIMit:SEGMent<Seg>:STIMulus:STOP
<FreqPowTime>
These commands change the start and stop stimulus values (i.e. the smallest and the
largest stimulus values) of a limit line segment. A segment must be created first to
enable the commands (e.g
ter 4.4.1.1, "Rules for Limit Line Definition"
Tip:
To define the stimulus values of several limit line segments with a single com-
CALCulate<Chn>:LIMit:CONTrol[:DATA]
.
SCPI Command Reference