The setting is identical with the "Selectivity" setting on the "Bandwidth" tab of the
"Power Bw Avg" softtool; see
Chapter 5.2.2, "Bandwidth Tab"
Remote command:
[SENSe<Ch>:]BWIDth[:RESolution]:SELect
5.5.4.4
Frequencies and Power Dialog
The"Frequencies and Power" dialog defines the (frequency or power) sweep ranges
for the lower tone. The frequency of the upper tone is equal to the lower tone fre-
quency plus the "Tone Distance" defined in the "Define Intermodulation Distortion Mea-
surement" dialog. Its power is equal to the lower tone power.
Access:
CHANNEL – CHANNEL CONFIG > "Intermod" > "Frequencies Powers"
Start Frequency / Stop Frequency / Power CW Frequency
The sweep range settings depend on the active sweep type (frequency or power
sweep). The frequency and power settings for the lower tone are identical with the
"Stimulus" settings, see
Chapter 5.8, "Stimulus Softtool"
Remote command:
[SENSe<Ch>:]FREQuency:STARt
[SENSe<Ch>:]FREQuency:STOP
[SENSe<Ch>:]FREQuency[:CW]
SOURce<Ch>:POWer<PhyPt>:STARt
SOURce<Ch>:POWer<PhyPt>:STOP
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate][:AMPLitude]
Frequency Range
The "Frequency Range" table shows the frequency ranges for the lower and upper
tone and for the highest-order measured IM product, depending on the current fre-
quency settings. E.g. the lower IM3 product is measured at the lower tone frequency
minus the tone difference. An red background in the "Range" column indicates that the
frequency range exceeds the upper or lower frequency limit of the analyzer.