active, a second receiver power calibration is performed at the converted lower
tone frequency.
If a free analyzer or generator port is used for the receiver power calibration (see
on page 384), an additional source power calibration for this port is
required. The additional calibration step is performed prior to step 3; the external power
meter must be connected to the calibrated port.
The progress of the calibration is monitored in the calibration sweep diagram, in close
analogy to an ordinary power sweep; see
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection:IMODulation:LTONe[:ACQuire]
SOURce<Ch>:POWer<PhyPt>:CORRection:IMODulation:UTONe[:ACQuire]
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
[SENSe<Ch>:]CORRection:POWer<PhyPt>:IMODulation:ACQuire
Apply
Applies the power calibration results and closes the calibration wizard.
5.5.5 Pwr Cal Settings Tab
Replicates the "Pwr Cal Settings" tab of the "Cal" softtool (see
5.5.6 Mixer Mode Tab
Controls the measurement of the mixing products generated by mixer under test (MUT)
that is supplied with two independent RF signals. The scalar mixer mode is included in
option R&S
ZNB/ZNBT-K4. An extension to a two-stage mixer system is possible.
Two dedicated SMARTerCal-Wizards, that are customized for the mixer measurement,
facilitate the channel calibration.
Background information
Refer to the following sections
●
Chapter 4.7.3.2, "Scalar Mixer Measurements"
●
"Two-Stage Mixer Measurements"
The buttons in the "SMARTerCal" section are enabled only if the mixer measurement
is set up properly. This is also required for the following reset functionality:
Channel Config Softtool