GUI Reference
R&S
®
ZNB/ZNBT
384
User Manual 1173.9163.02 ─ 38
Lower Tone
Selects an analyzer port as a source of the lower tone signal.
Remote command:
[SENSe<Ch>:]FREQuency:IMODulation:LTONe
Upper Tone
Selects an analyzer port or external generator as a source of the upper tone signal.
The source must be different from the lower tone source. This means that you cannot
combine coupled generator ports for the lower and upper tones (e.g. port 3/4 or port
1/2 on a four-port R&S
. If no internal second source
is available, the upper tone must be provided by a (previously configured) external
generator.
Note:
are part of the RF connection configuration, opera-
tion with internal second source is
not
supported. In this case only external generators
can be used as upper tone generator.
Remote command:
[SENSe<Ch>:]FREQuency:IMODulation:UTONe
Receiving Port
Selects the receiving port for the signal from the output of the DUT.
Remote command:
[SENSe<Ch>:]FREQuency:IMODulation:RECeiver
Src. for Rcvr. Cal
Selects the signal source for power calibration of the receiving port. In general, the
combiner output signal is most convenient and provides the simplest calibration proce-
dure.