GUI Reference
R&S
®
ZNB/ZNBT
549
User Manual 1173.9163.02 ─ 38
5.14.8 Set by Marker Tab
The "Set by Marker" functions use the active marker to define the sweep range, scale
the diagram and introduce an electrical length offset. The functions are unavailable if
the active trace contains no markers (e.g. after "All Markers Off").
Examples
Refer to
Chapter 3.3.6.6, "Set by Marker"
Center = Marker / Start = Marker / Stop = Marker / Span = Marker
The following functions use the stimulus value of the active marker to define the sweep
range.
●
"Center = Marker" sets the center of the sweep range equal to the stimulus value of
the active marker, leaving the span unchanged. The active marker appears in the
center of the diagram.
●
"Start = Marker" sets the beginning (start) of the sweep range equal to the stimulus
value of the active marker, leaving the end (stop) value unchanged. The active
marker appears at the left edge of the diagram.
●
"Stop = Marker" sets the end (stop) of the sweep range equal to the stimulus value
of the active marker, leaving the beginning (start) unchanged. The active marker
appears at the right edge of the diagram.
●
"Span = Marker" is only available if the active marker is in
span of the sweep range equal to the absolute value of the first coordinate of the
active marker, i.e. to the difference between the delta marker and the reference
marker positions. The reference marker appears at the right, the delta marker at
the left edge of the diagram or vice versa, depending on which of the two markers
has the smaller stimulus value.
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:CENTer
CALCulate<Chn>:MARKer<Mk>:FUNCtion:STARt
CALCulate<Chn>:MARKer<Mk>:FUNCtion:STOP
CALCulate<Chn>:MARKer<Mk>:FUNCtion:SPAN
Marker Softtool