Command Reference
R&S
®
ZNB/ZNBT
1145
User Manual 1173.9163.02 ─ 38
If several cal units are USB-connected to the analyzer, the command queries the cal
unit selected via
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
Query parameters:
<CalKitFileName>
String parameter containing the name of a cal unit characteriza-
tion.
'Factory'
denotes the factory characterization; an empty
string
' '
denotes the last referenced characterization; the pre-
fix "SD:" indicates that the characterization is stored on an SD
card inserted at the calibration unit.
Example:
See
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:
Usage:
Query only
Manual operation:
See
SYSTem:COMMunicate:RDEVice:AKAL:PORTs?
<CalKitFileName>
Queries the number of ports of a cal unit characterization (cal kit file)
<CalUnitName>
, the assigned connector types, and their gender. A possible
response for a two-port calibration is
'1,N 50 Ohm,MALE,2,N 50 Ohm,MALE'
. An
empty string is returned if no calibration unit is connected.
If several cal units are USB-connected to the analyzer, the command queries the cal
unit selected via
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
Query parameters:
<CalKitFileName>
String parameter containing the name of a cal unit characteriza-
tion.
'Factory'
denotes the factory characterization; an empty
string
' '
denotes the last referenced characterization; the pre-
fix "SD:" indicates that the characterization is stored on an SD
card inserted at the calibration unit.
Example:
See
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:
Usage:
Query only
Manual operation:
See
SYSTem:COMMunicate:RDEVice:AKAL:PREDuction[:STATe]
<Boolean>
Enables or disables automatic power reduction at all test ports while an automatic cali-
bration using the calibration unit R&S ZV-Zxx is active.
Parameters:
<Boolean>
Power reduction enabled or disabled.
*RST:
ON
SCPI Command Reference